Crome/Cobalt Respiration Study

단계 해당 없음 완료 NCT06089694

대상 질환Heart Failure
임상 단계단계 해당 없음
스폰서Medtronic Cardiac Rhythm and Heart Failure
상태완료
참여자 수61 명
중재(약물)Enabling Sensor Research Holter Mode feature
시작일2023년 11월 21일
완료일2024년 6월 12일
결과 공개일2026년 7월 21일

요약

The goal of this study is to compare measured respiration rates in various subject scenarios and use conditions between capnography and the Crome and Cobalt ICD and CRT-D MRI SureScan devices (derived using a Holter recorder).

Daily Capnography Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects15.3 ± 3.7 breaths per minute

Daily Device-derived Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects15.5 ± 3.5 breaths per minute

Difference in Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects0.2 ± 1.1 breaths per minute