Crome/Cobalt Respiration Study

該当なし 完了 NCT06089694

対象疾患Heart Failure
該当なし
スポンサーMedtronic Cardiac Rhythm and Heart Failure
ステータス完了
参加者数61 名
介入(薬剤)Enabling Sensor Research Holter Mode feature
開始日2023年11月21日
完了日2024年6月12日
結果公開日2026年7月21日

概要

The goal of this study is to compare measured respiration rates in various subject scenarios and use conditions between capnography and the Crome and Cobalt ICD and CRT-D MRI SureScan devices (derived using a Holter recorder).

Daily Capnography Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects15.3 ± 3.7 breaths per minute

Daily Device-derived Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects15.5 ± 3.5 breaths per minute

Difference in Respiration Rate

During In-clinic Testing Visit, an average of 2 hours · breaths per minute

All Enrolled Subjects0.2 ± 1.1 breaths per minute